Right Implant - patch without evidence of a gel fill hole or seal; patch markings "PIP 285cc 30901 050" appear contiguously under one another. The shell diameter is 126-130 mm. The patch measures 52-53 mm and is placed internally at a shell aperture of 36-37 mm. The approximate projection is 25-27 mm.
Left implant - patch without evidence of a gel fill hole or seal; patch markings "PIP 285cc 30901 056" appear contiguously under one another. The shell diameter is 126-130 mm. The patch measures 52-53 mm and is placed internally at a shell aperture of 36-37 mm. The approximate projection is 25-27 mm.
Observed shell dimenslons for both implants and residual weight correlate with PIP catalogues issued between 1998 and 2002. ., Implant Identification:' The implants are identified by direct inspection as Poly Implant Protheses (PIP) round, textured, "Profil Standard" (low profile) silicone gel-filled mammary prostheses nominally rated at 285 cc, on the basis of the above noted observations.